Product Details:
Payment & Shipping Terms:
|
Model: | SRX-Z75-T | Rated Voltage: | 75KVDC |
---|---|---|---|
Conductor Material: | Copper | Insulation Material: | Rubber |
High Light: | medical grade cable,insulated hv cable |
90º Plug 75KV HV Generator Accessories X-ray High Voltage Cables SRX-Z75-T
Description:
This 3-conductor with 90º Plug high voltage cable’s typical applications are as followings:
1, Medical x-ray equipment like standard x-ray, computer tomography and angiography equipment.
2, Industrial and scientific x-ray or electron beam equipment like electron microscopy and x-ray diffraction equipment.
3, Low power high voltage testing and measuring equipment
Technical Data
Number of conductor | 3 |
Rated voltage | 75kVDC |
Routine test voltage (high voltage insulation) | 120kVDC/10min |
Routine test voltage (conductor insulation) | 2kVACrms/1min |
Maximum conductor current | 1.5mm2:15A |
Nominal outside diameter | 17.0±0.5mm |
Thickness of PVC jacket | 1.0mm |
Thickness of high voltage insulation | 4.5mm |
Diameter of core-assembly | 4.5mm |
Insulation resistance core to shield @20℃ | ≥1×1012Ω·m |
Conductor insulation resistance@20℃ | ≥1×1012Ω·m |
Max Conductor resistance bare cond.@20℃ | 10.5mΩ/m |
Max Conductor resistance insul. cond. @20℃ | 12.2 mΩ/m |
Max Shield resistance@20℃ | 15 .0mΩ/m |
Max Capacitance between conductor and shield | 165nF/km |
Max Capacitance between ins. cond. and bare cord | 344nF/km |
Max Capacitance between insulated conductors | 300nF/km |
Cable Min bending radius(static insulation) | 40mm |
Cable Min bending radius(dynamic installation) | 80mm |
Operating temperature | -10℃~+70℃ |
Storage temperature | -40℃~+70℃ |
Net weight | 351kg/km |
Connection Diagram:
Cable Schema
Cable Assembly:
Photo of HV Cable Assembly