| Model: | SRMMWHX7360 | Nominal Tube Voltage: | 150KV |
|---|---|---|---|
| Nominal Focal Spot: | 0.6/1.2 | Exposure Rating: | 30KW/74KW(150/180Hz |
| Target Angle: | 12° | Rotation Speed: | 2800RPM/8400RPM |
| Anode Heat Storage Capacity: | 212 KJ (300kHU) | Max. Continuous Heat Dissipation: | 180W |
| Equivalent To: | Toshiba E7252/RAD14/RAY-14 | Warranty: | 12 Months |
| High Light: | dental x ray machine,portable x ray machine |
||
Rotating Anode X‐Ray tube assembly for medical x ray machine
Description
.● The designation is composed as follows:
| SRMWHX7360 | Tube | A | High voltage socket with 90 degree direction |
| SRMWTX73-0.6/1.2-150H | B | High voltage socket with 270 degree direction |
Technical Data
| Property | Specification | Standard | |
| Nominal input power(s) of the anode | F 1 | F 2 | IEC 60613 |
|
20kW(50/60Hz) 30kW(150/180Hz) |
50kW(50/60Hz) 74kW(150/180Hz) |
||
| Anode heat storage capacity | 212 k (300kHU) | IEC 60613 | |
| Maximum cooling capacity of the anode | 750W | ||
| Heat storage capacity | 900kJ | ||
|
Max. continuous heat dissipation without Air-circular |
180W | ||
|
Anode material Anode top coating material |
Rhenium-Tungsten-TZM(RTM) Rhenium-Tungsten-(RT) |
||
| Target angle (Ref: reference axis) | 12 ° | IEC 60788 | |
| X-ray tube assembly inherent filtration | 1.5 mm Al / 75kV | IEC 60601-1-3 | |
| Focal spot nominal value(s) | F1(small focus) | F2(large focus) | IEC 60336 |
| 0.6 | 1.2 | ||
|
X-ray tube nominal voltage Radiographic Fluoroscopic |
150kV 125kV |
IEC 60613 | |
| Max. Tube Current | Large Focus | 1200mA | |
| Small Focus | 500mA | ||
|
Data on cathode heating
Max. current Max voltage |
≈ /AC, < 20 kHz | ||
| F1 | F 2 | ||
|
5.4 A ≈9 V |
5.4 A ≈17 V |
||
|
Leakage radiation at 150 kV / 3mA in 1m distance |
≤ 0.5 mGy/h | IEC60601-1-3 | |
| Maximum radiation field | 430×430mm at SID 1m | ||
| X-ray tube assembly weight | Approx. 18 kg | ||
Stator key values
1-phase stator
| Test point | C-P | C-S |
| Winding resistance | ≈18.0…22.0Ω | ≈45.0…55.0Ω |
| Max. permissible operating voltage(run-up) | 230V±10% | |
| Recommend operating voltage(run-up) | 160V±10% | |
| Braking voltage | 70VDC | |
| Run-on voltage in exposure | 80Vrms | |
| Run-on voltage in fluoroscopy | 20V-40Vrms | |
| Run-up time(depending on starter system) | 1.2s | |
More Detailed Images
![]()
More product:
| Type | Focal spot | max.rating(KW) | Max voltag(kV) | Target angle() | Target material | Rotation rate(W/min) | Max component heat content(KJ) | Max anode heat content(KHU) | Max anode continuous heat dissipation(W) |
| SRMWHX6410 | 0.3/0.6 | 5/17(50Hz) | 130 | 10 | RTM | 2800 | 900 | 210 | 300 |
| SRMWHX6411 | 0.6/1.3 | 11/32(50Hz) | 130 | 15 | RTM | 2800 | 900 | 110 | 300 |
| SRMWHX6412 | 0.8/1.8 | 12/33(50Hz) | 130 | 15 | RTM | 2800 | 900 | 70 | 175 |
| SRMWHX6460 | 0.6/1.3 | 11/33(50Hz) | 130 | 15 | RTM | 2800/8400 | 900 | 110 | 300 |
| 20/54(150Hz) | |||||||||
| SRMWHX7010 | 1.0/2.0 | 21/42.5(50Hz) | 125 | 16 | RTM | 2800 | 900 | 140 | 475 |
| SRMWHX7110 | 0.6/1.2 | 20/40(50Hz | 125 | 12.5 | RTM | 2800 | 900 | 150 | 500 |
| SRMWHX7210 | 1.0/2.0 | 20/40(50Hz) | 125 | 16 | TM | 2800 | 900 | 140 | 410 |
| SRMWHX7220 | 1.0/2.0 | 20/40(50Hz) | 150 | 16 | TM | 2800 | 900 | 140 | 410 |
| SRMWHX7360 | 0.6/1.2 | 20/50(50Hz) | 150 | 12 | RTM | 2800/8400 | 900 | 300 | 750 |
| 30/74(150Hz) | |||||||||
| SRMWHX7420 | 0.6/1.2 | 20/46(50Hz) | 150 | 12 | RTM | 2800 | 900 | 150 | 475 |
| SRMWHX7520 | 0.6/1.2 | 20/50(50Hz) | 150 | 12 | RTM | 2800 | 900 | 220 | 750 |
| SRMWHX7620 | 0.6/1.2 | 20/50(50Hz) | 150 | 12 | RTM | 2800 | 900 | 300 | 870 |
| SRMWHX7810 | 0.6/1.5 | 18/47.5(50Hz) | 125 | 14 | RTM | 2800 | 900 | 200 | 410 |